Installing the GC
1
Agilent Intuvo 9000 GC Installation
15
Verify line voltage and power cord.
1
Verify that the power cord is correct for the voltage and
location. See
The power cord termination is determined by the country
where the GC is ordered.
2
Have a qualified person measure the actual power outlet
voltage and verify it meets the tolerance requirements listed
in
The next sections detail the power specifications and
requirements for reference.
Power consumption
The number and type of electrical outlets required for
installation depends on the size and complexity of your system.
A GC system with a computer, monitor, printer, and hub
requires five outlets.
The GC does not require a dedicated
ground, but since the GC can draw up to 12 A maximum,
depending on the line voltage, you may not want other
devices on the same circuit.
Power line conditioners should not be used with the GC.
Power cords available
lists the power cords available for the GC. If your power
cord is incorrect, order the cord appropriate for the country.
WARNING
Electrical shock hazard. To avoid injury, only a qualified person
should measure line voltage.
Table 1
GC power requirements
Line voltage
*
Frequency
Current
Maximum continuous
power consumption (VA)
Power outlet current
rating
120 single phase*(–10% / +10%)
50/60
(–5% / +5%)
12 A
1296
15 A
200–240 V single/split phase,
+10 to –10%
50/60
(–5% / +5%)
7.7 / 6.5
1548
10 A
* Users of 100 V power (Japan) receive the 120 V US power option with a 100 V/120 V step up transformer.
Summary of Contents for Intuvo 9000
Page 1: ...Agilent Technologies Agilent Intuvo 9000 Gas Chromatograph Installation and First Startup...
Page 6: ...6 Agilent Intuvo 9000 GC Installation...
Page 25: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 25 3 Reinstall the GC covers...
Page 27: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 27 Turn on the GC...
Page 29: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 29...
Page 84: ...84 Agilent Intuvo 9000 GC Installation 1 Installing the GC 8 Close the GC front door...
Page 114: ...Agilent Technologies...