36
Agilent Intuvo 9000 GC Installation
1
Installing the GC
Prepare the GC.
1
Remove protective packing used for shipment.
•
On the back of the GC, remove the caps from the inlet and
detector gas fittings.
•
Remove any cap from the cryogenic coolant fitting, if
present.
•
Remove any protective cap from the detector exhaust port
or exhaust tubing.
•
On the top of the GC, remove any tape used to secure
covers, and remove any protective caps placed on detector
exhaust port.
Back panel caps
Tape and detector exhaust cap
Door clamp
T20
Cryo coolant fitting cap
(if present)