Cabling Diagrams and Remote Start/Stop
B
Agilent Intuvo 9000 GC Installation
113
When used for external event control, apply label G1580-87200
to identify the cable for EVENT use.
External valve cable, G1580-60710
Supplies power for certain valve applications.
7
Closure 2
48 V AC/DC, 250 mA
Brown or violet
8
8
Closure 2
Blue
8
Table 13
External events cable (continued)
Connector 1 pin
Signal name
Maximum rating
Connector 2, wire colorControlled by valve #
Connector 1 pin Wire color
Connector and pin
Function
1
Yellow
V5 pin 1
24 V, 150 mA max.
2
Black
V6 pin 1
24 V, 150 mA max.
3
Red
V5 pin 2
Ground
4
White
V6 pin 2
Ground
5
6
7
8
2
8
1
6
2
1
2
1
EVENT
CAUTION: DO NOT PLUG
INTO THE BCD PORT
.
V5
V6
Summary of Contents for Intuvo 9000