Reference
8
6000 Series Oscilloscope User’s Guide
341
Figure 49
DC and Low-Frequency Probe Equivalent Circuit
Figure 50
High-Frequency Probe Equivalent Circuit
The impedance plots for the two models are shown in these
figures. By comparing the two plots, you can see that both the
series tip resistor and the cable’s characteristic impedance
extend the input impedance significantly. The stray tip
capacitance, which is generally small (1 pF), sets the final break
point on the impedance chart.
8.5 pF
100 k
W
150
W
1 pF
7.5 pF
100 k
W
250
W
Summary of Contents for InfiniiVision 6000 Series