Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
159
In Asynchronous Mode
1
Press the
Baud
softkey to select a baud rate of
2.5 Mb/s
,
5 Mb/s
, or
10 Mb/s
.
2
Press the
Channel
softkey to select channel
A
or
B
.
In Synchronous Mode
1
Press the
FIBEX File
softkey to specify the location of the
FIBEX (Field Bus Exchange format) file that defines the
time-triggered timing schedule of the probed FlexRay bus.
2
Press the
Channel
softkey to select channel
A
or
B
.
In Asynchronous or Synchronous Mode
In addition to the logic
probe cable connection between the VPT1000 and the
oscilloscope, you must make a LAN connection to allow the
oscilloscope to control the VPT1000. See also
“Specifying the
VPT1000 LAN Address”
on page 157.
N O T E
If the VPT1000 loses synchronization, then press the
Mode
key (in the
VPT1000 Menu) on the oscilloscope to cycle through the modes and
restore synchronization.
N O T E
Before powering-off the oscilloscope, change the VPT1000 2 mode to
PC
.
If oscilloscope power is cycled when in Asynchronous or Synchronous
modes, communication with the VPT1000 will be lost. To restore
communication with the VPT1000, cycle the VPT1000 power, wait for its
State indicator to turn from red to green, and then press the
Mode
key (in
the VPT1000 Menu) on the oscilloscope to cycle through the modes.
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...