Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
147
Hex Bus Pattern Triggering
You can specify a bus value on which to trigger. To do this, first
define the bus. See
“To display digital channels as a bus”
on
page 122 for details. You can trigger on a bus value whether you
are displaying the bus or not.
To trigger on a bus value:
1
Press the
Pattern
key on the front panel.
2
Press the
Channel
softkey and rotate the Entry knob to select
Bus1
or
Bus2
.
3
Press the
Digit
softkey and rotate the Entry knob to select a
digit of the selected bus.
4
Press the
Hex
softkey and rotate the Entry knob to select a
value for the digit.
5
You can use the
Set all Digits
softkey to set all digits to a
particular value.
When a hex bus digit contains one or more don’t care (X) bits
and one or more bit with a value or 0 or 1, the “$” sign will be
displayed for the digit.
For information regarding digital bus display when Pattern
triggering see
“Bus values are displayed when using Pattern
trigger”
on page 125.
N O T E
If a digit is made up of less than four bits, then the value of the digit will be
limited to the value that can be created by the selected bits.
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...