104
6000 Series Oscilloscope User’s Guide
2
Front-Panel Controls
3
Turn the Entry knob to select the number of minutes to wait
before the selected screen saver activates.
When you turn the Entry knob, the number of minutes is
displayed in the
Wait
softkey. The default time is 180 minutes
(3 hours).
4
Press the
Preview
softkey to preview the screen saver you
have selected with the
Saver
softkey.
5
To view the normal display after the screen saver has started,
press any key or turn any knob.
To set the waveform expansion reference point
•
Press
Utility
&
Options
&
Preferences
&
Expand
and select
Ground
or
Center
.
When you change a channel's volts/division setting, the
waveform display can be set to expand (or compress) about the
signal ground level or the center of the display.
Expand About Ground
The displayed waveform will expand
about the position of the channel's ground. This is the default
setting. The ground level of the signal is identified by the
position of the ground level (
) icon at the far-left side of the
display. The ground level will not move when you adjust the
vertical sensitivity (volts/division) control.
If the ground level is off screen, the waveform will expand about
the top or bottom edge of the screen based on where the ground
is off screen.
Expand About Center
The displayed waveform will expand
about the center of the display.
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...