Making Measurements
5
6000 Series Oscilloscope User’s Guide
215
Multiply
When you select
1 * 2
, channel 1 and channel 2 voltage values
are multiplied point by point, and the result is displayed. 1 * 2 is
useful for seeing power relationships when one of the channels
is proportional to the current.
1
Press the
Math
key, press the
1 * 2
softkey, then press the
Scale
or
Offset
sofkeys if you want to change the scaling or offset for
the multiply function.
•
Scale
— lets you set your own vertical scale factors for
multiply expressed as V
2
/div (Volts-squared/division),
A
2
/div (Amps-squared/division), or W/div (Watts/division
or Volt-Amps/division). Units are set in the channel
Probe
menu. Press the
Scale
softkey, then turn the Entry knob to
rescale 1 * 2.
•
Offset
— lets you set your own offset for the multiply math
function. The offset value is in V
2
(Volts-squared), A
2
(Amps-squared), or W (Watts) and is represented by the
center horizontal grid line of the display. Press the
Offset
softkey, then turn the Entry knob to change the offset for
1 * 2.
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...