6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
For example, for a positive pulse, if you set t>10 ns:
•
within a range of time values (
><
).
For example, for a positive pulse, if you set t>10 ns and
t<15 ns:
5
Select the qualifier time set softkey (
<
or
>
), then rotate the
Entry knob to set the pulse width qualifier time.
The qualifiers can be set as follows:
•
2 ns to 10 s for
>
or
<
qualifier (5 ns to 10 s for 100 MHz
and 300 MHz bandwidth models)
•
10 ns to 10 s for
><
qualifier, with minimum difference of
5 ns between upper and lower settings
< qualifier time set softkey
•
When the less than (
<
) qualifier is selected, the Entry knob
sets the oscilloscope to trigger on a pulse width less than the
time value displayed on the softkey.
•
When the time range (
><
) qualifier is selected, the Entry
knob sets the upper time range value.
> qualifier time set softkey
•
When the greater than (
>
) qualifier is selected, the Entry
knob sets the oscilloscope to trigger on a pulse width greater
than the time value displayed on the softkey.
•
When the time range (
><
) qualifier is selected, the Entry
knob sets the lower time range value.
Trigger
10 ns
10 ns
Trigger
12 ns
15 ns
10 ns
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...