174
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
4
Press the
Signals
softkey. The LIN Signals Menu will be
displayed.
5
Press the
Source
softkey to select the channel connected to
the LIN signal line.
Adjust the trigger level for the selected analog channel by
turning the Trigger Level knob. Press the
D15 Thru D0
key and
select
Thresholds
to set the threshold level for digital
channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
6
Press the
Baud
softkey to set the LIN signal baud rate to
match your LIN bus signal.
The LIN baud rate can be set to 2400 b/s, 9600 b/s, 10.4 kb/s,
19.2 kb/s, 115.2 kb/s, or 625 kb/s. The default baud rate is
2400 b/s.
7
Press the
Smpl Pt
softkey to select the sample point at which
the oscilloscope will sample the bit value.
8
Press the
Standard
softkey to select the LIN standard your are
measuring (LIN 1.3 or LIN 2.0).
9
Press the
Sync Break
softkey and select the minimum number
of clocks that define a sync break in your LIN signal.
Signal
baud rate
Source
Menu
Return to
previous menu
Sample
Point
Define Sync
Break
Standard
Selector
One Bit
60%
70%
80%
Sample Point
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...