Features and Functions
2
34980A User’s Guide
57
•
To ensure you are getting the most accurate measurements possible,
use care when setting the channel delay less than the default value
(automatic). The default channel delay is designed to optimize
parameters, such as settling time, for the most accurate measurements.
•
The
CONFigure
and
MEASure?
commands set the channel delay to
automatic. A Factory Reset (
*RST
command) also sets the channel delay
to automatic.
Front Panel Operation:
Channel (Configure) > CHANNEL DELAY > TIME
Once you have added the specified channel to the scan list, the channel
delay choice will be visible in the menu.
Interface Operation:
The following command add a 2- second channel delay
to the specified channels.
ROUT:CHAN:DELAY 2,(@1003,1013)
Automatic Channel Delays
If you do not specify a channel delay, the instrument selects a delay
for you. The delay is determined by the delay based on function, range,
integration time, and ac filter setting.
DC Voltage, Thermocouple, DC Current (for all ranges):
Resistance, RTD, Thermistor (2- and 4-wire):
Integration Time
Channel Delay
PLC > 1
PLC
≤
1
2.0 ms
1.0 ms
Range
Channel Delay
(for PLC > 1)
Range
Channel Delay
(for PLC
≤
1)
100
Ω
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
10 M
Ω
100 M
Ω
2.0 ms
2.0 ms
2.0 ms
25 ms
30 ms
200 ms
200 ms
100
Ω
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
10 M
Ω
100 M
Ω
1.0 ms
1.0 ms
1.0 ms
20 ms
25 ms
200 ms
200 ms
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...