258
34980A User’s Guide
8
Microwave Switch/Attenuator Driver
SCPI Programming Examples
These programming examples provide you with SCPI command examples
to use for driving the microwave switch modules.
The slot and channel addressing scheme used in these examples follow the
form
srcc
where
s
is the mainframe slot number (1 through 8),
r
is the
remote module number (1 through 8), and
cc
is the two- digit channel
number. For more information about channel numbering, refer to
“Channel
Numbering”
on page 212.
For complete information on SCPI commands, see the
Programmer’s
Reference
Help file.
Example: Configuring an Agilent N1810UL
The following example illustrate controlling an Agilent N1810UL attached
to a Y1150A distribution board. The distribution board is connected to
Bank 1 of the first remote module attached to the 34945A installed in
slot 1 of the mainframe. This example uses the bank preset (described on
page 223
).
ROUTe:RMODule:DRIVe:SOURce OFF,(@1100)
ROUTe:RMODule:BANK:PRESet BANK1,(@1100)
ROUTe:RMODule:DRIVe:SOURce INT,(@1101)
ROUT:CLOSe (@1101)
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...