Features and Functions
2
34980A User’s Guide
59
Reading Format
During a scan, the instrument automatically adds a time stamp to all
readings and stores them in memory. Each reading is stored with
measurement units, time stamp, channel number, and alarm status
information. From the remote interface, you can specify which information
you want returned with the readings (from the front panel, all of the
information is available for viewing). The examples below show a reading
in
relative
and
absolute
format with all fields enabled.
Relative Format (Default):
Absolute Format:
•
The reading format applies to all readings being removed from the
instrument from a scan; you cannot set the format on a per- channel
basis.
•
The
CONFigure
and
MEASure?
commands automatically turn off the
units, time, channel, and alarm information.
•
The format settings are stored in volatile memory and will be lost when
power is turned off or after a Factory Reset (
*RST
command).
1
2
Reading with units (26.195 °C)
Time since start of scan (17 ms)
3
4
Channel number
Alarm limit threshold crossed
(0 = No Alarm, 1 = LO, 2 = HI)
1
2
3
Reading with units (26.195 °C)
Date (November 21, 2004)
Time of day (3:30:23.000 PM)
4
5
Channel number
Alarm limit threshold crossed
(0 = No Alarm, 1 = LO, 2 = HI)
2.6101 C,000000000.017,1003,2
1
2
4
3
2.6101 C,2004,11,21,15,30,23.000,1003,2
1
3
5
4
2
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...