268
34980A User’s Guide
10
64-Bit Digital I/O Module with Memory and Counter
To read digital data with more control over the channel parameters,
use the SCPI
CONFigure
and
SENSe
commands. The
CONFigure
commands
set up the digital I/O channel parameters. For example, sending the
following SCPI command to a Digital I/O module installed in slot 1 of the
mainframe, sets a 16- bit input channel (103) to use a 2.5 V input
threshold, and normal polarity.
CONF:DIG WORD, 2.5, NORM, (@1103)
Once configured, the data is read using the following command.
SENS:DIG:DATA:WORD? (@1103)
You may also read an individual bit using the
SENSe
commands.
This allows you to check the state of an individual bit in a channel
without having to create an input mask. For example, the following
command returns the state of bit 3 in the channel 101 byte.
SENS:DIG:DATA:BIT? 3, (@1101)
The acceptable range for the bit parameter is based on the channel width
as shown below:
•
BYTE (8- bit): <bit> can range from ‘0’ to ‘7’
•
WORD (16- bit): <bit> can range from ‘0’ to ‘15’
•
LWORd (32- bit): <bit> can range from ‘0’ to ‘31’
The
SENSe
command differs from the
MEASure
command in that it will not
change the direction (input or output) of the channel. If the channel is
configured as an output, the
SENSe
command will return the value being
driven.
Writing Digital Data
To write digital data, set the channel output parameters using the
SOURce
commands. For example, sending the following SCPI commands to a Digital
I/O module in slot 1 sets a 32- bit channel to use normal polarity,
with active drive and a ‘set’ output voltage of 4 volts.
CONF:DIG:WIDT LWOR,(@1201)
CONF:DIG:POL NORM,(@1201)
SOUR:DIG:DRIV ACT,(@1201)
SOUR:DIG:LEV 4,(@1201)
The width and polarity parameters apply to both input and output
operations.
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...