Features and Functions
2
34980A User’s Guide
89
Front Panel Operation:
Store/Recall > STORE|RECALL|DELETE|RENAME|AUTO
To rename a location, select
RENAME
. Press the arrow keys to move the
cursor to a specific position and then turn the knob to select the desired
letter or number. To clear the name of a location, change each character
to “
^
” (starting with the rightmost character) and then press the left
arrow key to move to the next character.
To automatically recall a specific location when power is restored,
select
AUTO
. Use the knob to scroll through the available locations
containing a stored state.
Remote Interface Operation:
Use the following commands to store and
recall instrument states.
*SAV {1|2|3|4|5}
*RCL {1|2|3|4|5}
To assign a user- defined name to a stored state to be recalled from the
front panel, see the following example. From the remote interface, you can
only recall a stored state using a number (1 through 5).
MEM:STAT:NAME 1,TEST_RACK_1
To configure the instrument to automatically recall location 2 when power
is restored, send the following commands.
*SAV 2
MEM:STATE:RECALL:SELECT 2
MEM:STATE:RECALL:AUTO ON
Error Conditions
When the front panel ERROR annunciator turns on, one or more command
syntax or hardware errors have been detected. A record of up to 20 errors
can be stored in the instrument’s
error queue
. Each remote interface I/O
session (i.e., GPIB, USB, LAN, etc.) has its own interface- specific error
queue. Errors appear in the error queue of the I/O session that caused
the error (the front panel reports errors from all I/O sessions).
For a complete listing of the error messages, see the Agilent 34980A
Programmer’s Reference
Help file, located on the
Product Reference
CD- ROM
shipped with the instrument.
•
The instrument beeps once each time a command syntax or hardware
error is generated.
•
A special
global
error queue holds all power- on and hardware- related
errors (e.g., over- temperature, Safety Interlock, etc.).
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...