Features and Functions
2
34980A User’s Guide
75
•
You can control the slope of the pulse from the alarm outputs
(the selected configuration is used for all four outputs). In the falling
edge mode, 0V (TTL low) indicates an alarm. In the rising edge mode,
+5V (TTL high) indicates an alarm. A Factory Reset (
*RST
command)
will reset the slope to falling edge.
Note
:
Changing the slope of the output lines may cause the lines to
change state.
Front Panel Operation:
•
To manually clear all four alarm output lines, select:
Alarm > CLEAR ALARM OUT? > YES|NO
•
To select the output configuration for all four output lines, select:
Alarm > ALARM OUT SIGNAL > TRACK|LATCH
•
To configure the slope of all four output lines, select:
Alarm > ALARM OUT SLOPE > NEGATIVE|POSITIVE
Remote Interface Operation:
To clear the specified output lines (or to clear
all four lines), use one of the following commands.
To select the output configuration for all four output lines, use the
following command.
OUTPut:ALARm:MODE {LATCh|TRACk}
To configure the slope of all four output lines, use the following command.
OUTPut:ALARm:SLOPe {NEGative|POSitive}
OUTPUT:ALARM2:CLEAR
OUTPUT:ALARM:CLEAR:ALL
Clear alarm output line 2
Clear all four alarm outputs
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...