138
34980A User’s Guide
4
Low Frequency Multiplexer Switch Modules
Four-Wire Mode
This 20- channel 4- wire MUX This configuration requires neither using
external wiring nor connecting through the internal Analog Buses.
For 4- wire resistance measurements, the instrument automatically pairs
channel
n
on Bank 1 with channel
n
+20 (Bank 2) to provide the source
and sense connections. Four- wire controls occur only when doing 4- wire
measurement operations through the internal DMM, such as
MEASure:FRESistance?
or scanning a channel previously configured as
4- wire.
One-Wire Mode
•
two independent 40- channel 1- wire MUXes. This configuration
requires neither using external wiring nor connecting through the
internal Analog Buses.
•
one 80- channel 1- wire MUX. You must use external wiring or connect
through the internal Analog Bus for this configuration.
In 1- wire mode, you can close no more than 40 channels simultaneously
due to power dissipation. These channels are split 20 to a bank. For
example, with one Analog Bus relay closed you can close up to a
maximum of 39 channel relays. If you try to close more than the allowed
number of channels, you will receive an error message.
In all modes, this module has capability to scan as many as 500
channels/second using the internal DMM. With the automatic
“break- before- make” connection operation, you are assured that no two
signals are connect to each other during a scan.
This module is interlock protected, which means whenever the D- sub
connector end of the modules is exposed, the Analog Bus relays
immediately open and disconnect from the Analog Bus. For more
information, refer to
page 120
, and
page 141
or
page 144
.
N O T E
Because all bank relays supply only HI signals, you can apply a
LOW signal through COM1 L or COM2 L when you are making
2-wire resistance measurements in 1-wire mode.
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...