Microwave Switch/Attenuator Driver
8
34980A User’s Guide
217
Using Pulse Drive
To use the pulse drive mode, send the
ROUTe:CHANnel:DRIVe:PULSe:MODE ON
command or pair two channels
with the
ROUTe:CHANnel:DRIVe:PAIRed:MODE
command. The diagram
below illustrates the pulse drive for two channels (switches) and the
relationship of the drive parameters to the power supply requirements.
As shown in the diagram, the drive is applied to channel 1 and held for
the T(Pulse) time set using the
ROUTe:CHANnel:DRIVe:PULSe:WIDTh
command. Drive is applied to channel 2 only after a power supply
recovery period has elapsed T(Recovery). The power supply recovery
time is set using the
ROUTe:CHANnel:DRIVe:TIMe:RECovery
command.
This parameter may be set individually for each channel or will default
to 0.0 ms following either a
SYSTem:RMODule:RESet
or
ROUTe:RMODule:BANK:PRESet
command.
If you are verifying the channel operation (see
page 218
), you may also
specify a T(Settle) parameter. During T(Settle) the switch is considered
‘busy’. This parameter ensures the switch has had time to change state
before the verification. This parameter may be set individually for each
channel or will default to 0.0 ms following either a
SYSTem:RMODule:RESet
or
ROUTe:RMODule:BANK:PRESet
command.
Unlike other switch modules, the 34945A will always pulse a channel in
response to a
ROUTe:OPEN
or
ROUTe:CLOSe
command. For example, sending
ROUTe:CLOSe
to a channel three times in a row will result in three
output pulses.
T(Setttle)
T(Recovery)
Drive Ch 1
Drive Ch2
Start Drive
Channel 1
Start Drive
Channel 2
Channel 1 Position
Indicators Evaluated
T(Pulse)
N O T E
A single drive channel operating in pulse mode with channel verification
(see
page 218
) turned off (default) will report the channel as ‘stateless’
and the
ROUTe:CLOSe?
query will return an error. Single drive pulsed
channels must have verification enabled
(
ROUTe:CHANnel:VERify ON
) to query the channel state using the
ROUTe:CLOSe?
query.
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...