Features and Functions
2
34980A User’s Guide
35
The following command sets the nominal resistance (R
0
) to 1000
Ω
on
channel 1003.
SENS:TEMP:TRAN:FRTD:RES 1000,(@1003)
Thermistor Measurements
The instrument supports 2.2 k
Ω
(YSI Series 44004), 5 k
Ω
(YSI Series 44007),
and 10 k
Ω
(YSI Series 44006) thermistors.
Front Panel Operation:
To select the thermistor function for the active
channel, choose the following items.
DMM or Channel (Configure) > TEMPERATURE > PROBE TYPE > THERMISTOR
To select the thermistor type for the active channel, choose from the
following items.
THERMISTOR TYPE > 10K|5K|2.2K
Remote Interface Operation:
You can use the
CONFigure
or
MEASure?
command to select the probe type and thermistor type. For example,
the following command configures channel 3001 for measurements of a
5 k
Ω
thermistor:
CONF:TEMP THER,5000,(@3001)
You can also use the
SENSe
command to select the probe type and
thermistor type. For example, the following command configures channel
1003 for measurements of a 10 k
Ω
thermistor:
SENS:TEMP:TRAN:THERM:TYPE 10000,(@1003)
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...