Low Frequency Multiplexer Switch Modules
4
34980A User’s Guide
129
34921A D-Sub Connectors
Bank 1
Bank 2
For orientation, the D-sub connector
end of the module is facing you.
Bank 2
Bank 1
17L
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
TSIL* 11H
11L
7H
7L
17H
13H
13L
9H
9L
19H
19L
15H 15L Interlock 1
Interlock1
1H
1L
2H
2L
3H
3L
COM
1H
COM
1L
4H
4L
14H
14L
5H
5L
20H
20L
18H
GND
6H
6L
16H
16L 12H
12L
8H
8L
18L
10H
10L
AMP
41L
AMP
41
I
AMP
42L
AMP
42
I
*TSIL represents
Temperature Sensor
Interface Line. This line
is used for temperature
interface only.
50-Pin D-Sub
Male Connector
Description Pin
Description Pin
Description Pin
Description Pin
Description Pin
1H
1
6H
35
11H
19
16H
37
COM1 H
7
1L
2
6L
36
11L
20
16L
38
COM1 L
8
2H
3
7H
21
12H
39
17H
23
Interlock 1
17
2L
4
7L
22
12L
40
17L
24
Interlock 1
33
3H
5
8H
41
13H
25
18H
43
GND
34
3L
6
8L
42
13L
26
18L
44
TSIL*
18
4H
9
9H
27
14H
11
19H
29
AMP 41L
47
4L
10
9L
28
14L
12
19L
30
AMP 41
I
48
5H
13
10H
45
15H
31
20H
15
AMP 42L
49
5L
14
10L
46
15L
32
20L
16
AMP 42
I
50
WARNING::
As a safety
feature, interlock 1 pins
(17 and 33) on Bank 1
must be shorted to
enable the Bank 1 Analog
Bus relays to close. The
optional 34921T terminal
block shorts these pins
for you. This feature
protects inadvertent
routing of high voltages
from the Analog Bus to
the D-sub connector of
the module.
WARNING
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
23L
Interlock 2
21H
21L
22H
22L 23H
COM
2H
COM
2L
24H
24L 34H
34L
25H
25L
40H
40L
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
37L
TSIL* 31H
31L
27H
27L 37H
33H
33L
29H
29L
39H
39L
35H 35L Interlock 2
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
GND
26H
26L 36H
36L 32H
32L
28H
28L 38H
38L
30H 30L
AMP
43L
AMP
43
I
AMP
44L
AMP
44
I
*TSIL represents
Temperature Sensor
Interface Line. This line
is used for temperature
interface only.
50-Pin D-Sub
Male Connector
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
21H
1
26H
35
31H
19
36H
37
COM2 H
7
21L
2
26L
36
31L
20
36L
38
COM2 L
8
22H
3
27H
21
32H
39
37H
23
Interlock 2
17
22L
4
27L
22
32L
40
37L
24
Interlock 2
33
23H
5
28H
41
33H
25
38H
43
GND
34
23L
6
28L
42
33L
26
38L
44
TSIL*
18
24H
9
29H
27
34H
11
39H
29
AMP 43L
47
24L
10
29L
28
34L
12
39L
30
AMP 43
I
48
25H
13
30H
45
35H
31
40H
15
AMP 44L
49
25L
14
30L
46
35L
32
40L
16
AMP 44
I
50
WARNING::
As a safety
feature, interlock 2 pins
(17 and 33) on Bank 2
must be shorted to
enable the Bank 2 Analog
Bus relays to close. The
optional 34921T terminal
block shorts these pins
for you. This feature
protects inadvertent
routing of high voltages
from the Analog Bus to
the D-sub connector of
the module.
WARNING
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...