34980A User’s Guide
2
Features and Functions
RTD Measurements
•
The instrument supports RTDs with
α
= 0.00385 (DIN/IEC 751) using
ITS- 90 software conversions or
α
= 0.00391 using IPTS- 68 software
conversions. The default is
α
= 0.00385.
•
The resistance of an RTD is nominal at 0 °C and is referred to as R
0
.
The instrument can measure RTDs with R
0
values from 49
Ω
to 2.1 k
Ω
.
•
You can measure RTDs using a 2- wire or 4- wire measurement method.
The 4- wire method provides the most accurate way to measure small
resistances. Connection lead resistance is automatically removed using
the 4- wire method.
•
For 4- wire RTD measurements, the instrument automatically pairs
channel
n
in Bank 1 with channel
n+20
in Bank 2 (34921A, 34923A) or
n+35
(34922A, 34924A) to provide the source and sense connections.
For example, make the source connections to the HI and LO terminals
on channel 2 in Bank 1 and the sense connections to the HI and LO
terminals on channel 22 (or 37) in Bank 2.
Front Panel Operation:
To select the 2- wire or 4- wire RTD function for the
active channel, choose the following items.
DMM or Channel (Configure) > TEMPERATURE > PROBE TYPE > RTD|4W RTD
To select the RTD type (
α
= 0.00385 or 0.00391) for the active channel,
choose the following item.
RTD TYPE > 0.00391|0.00385
To select the nominal resistance (R
0
) for the active channel, choose the
following item.
RO > 100 OHM
Remote Interface Operation:
You can use the
CONFigure
or
MEASure?
command to select the probe type and RTD type. For example, the
following command configures channel 3001 for 2- wire measurements of
an RTD with
α
= 0.00385 (use “85” to specify
α
= 0.00385 or “91” to
specify
α
= 0.00391).
CONF:TEMP RTD,85,(@3001)
You can also use the
SENSe
command to select the probe type, RTD type,
and nominal resistance. For example, the following command configures
channel 1003 for 4- wire measurements of an RTD with
α
= 0.00391
(channel 1003 is automatically paired with channel 1023 for the 4- wire
measurement).
SENS:TEMP:TRAN:FRTD:TYPE 91,(@1003)
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...