34980A User’s Guide
2
Features and Functions
Measurement Range
You can allow the instrument to automatically select the measurement
range using
autoranging
or you can select a fixed range using
manual
ranging
. Autoranging is convenient because the instrument decides which
range to use for each measurement based on the input signal. For fastest
scanning operation, use manual ranging on each measurement (some
additional time is required for autoranging since the instrument has to
make a range selection).
•
Autorange thresholds:
•
If the input signal is greater than can be measured on the selected
range (manual ranging), the instrument gives an overload indication:
“±OVLD” from the front panel or “±9.9E+37” from the remote interface.
•
For temperature measurements, the instrument internally selects the
range you cannot select which range is used. For thermocouple
measurements, the instrument internally selects the 100 mV range.
For thermistor and RTD measurements, the instrument autoranges to
the correct range for the transducer resistance measurement.
•
For frequency and period measurements, the instrument uses one
“range” for all inputs between 3 Hz and 300 kHz. The
range
parameter
is required only to specify the resolution. Therefore, it is not necessary
to send a new command for each new frequency to be measured.
•
The
CONFigure
and
MEASure?
commands contain an optional parameter
which allows you to specify the range or autoranging.
•
The instrument returns to autoranging when the measurement function
is changed and after a Factory Reset (
*RST
command). An Instrument
Preset (
SYSTem:PRESet
command) or Card Reset (
SYSTem:CPON
command) does
not
change the range setting.
Front Panel Operation:
DMM or Channel (Configure) > RANGE
First, select the measurement function on the active channel. You are
automatically guided to the next level of the menu where you can select
a specific range or autoranging.
Remote Interface Operation:
You can select the range using parameters in
the
CONFigure
and
MEASure?
commands. For example, the following
command selects the 10 Vdc range on the specified channel.
CONF:VOLT:DC 10,DEF,(@3001)
Down range at:
Up range at:
<10% of range
>120% of range
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...