66
34980A User’s Guide
2
Features and Functions
•
For an externally- controlled scan, you must either remove the internal
DMM from the 34980A or disable it (see
“Internal DMM Disable”
on
page 93). Since the 34980A’s internal DMM is not used, readings from
multiplexer channels are stored in the external DMM’s memory.
•
In this configuration, you must set up a
scan list
to include all desired
multiplexer or digital channels. Channels which are not in the list are
skipped during the scan. By default, the instrument scans the list of
channels in ascending order from slot 1 through slot 8 (channels are
reordered as needed).
•
You can configure the event or action that controls the onset of each
sweep through the scan list (a
sweep
is one pass through the scan list).
The selected source is used for all channels in the scan list. For more
information, refer to
“Scan Trigger Source”
on page 47.
•
You can configure the event or action that notifies the 34980A to
advance to the next channel in the scan list. Note that the
Channel
Advance
source shares the same sources as the scan trigger. However,
an error is generated if you attempt to set the channel advance source
to the same source (other than IMMediate) used for the scan trigger.
1
5
6
9
5
1
9
6
ABus1 HI
ABus2 HI
ABus3 HI
ABus4 HI
ABus1 LO
ABus2 LO
ABus3 LO
ABus4 LO
Chan Adv In
GND
Chan Closed Out
34980A
External DMM
External DMM
VM Complete Out
Ext Trig In
Analog Bus Connector
34980A
Ext Trig Connector
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...