50
34980A User’s Guide
2
Features and Functions
Front Panel Operation:
Scan (Configure) > ALARM
To enable the Monitor function, select the desired channel and then press
the
DMM
or
Channel (Measure)
key. To initiate the scan, press the
Scan
(Measure)
key. When an alarm occurs, the scan starts and readings are
stored in memory.
Note:
To stop a scan, press and hold the
Scan (Measure)
key.
You can also configure whether the instrument sweeps through the scan
list one time or continuously when an alarm condition is detected:
Scan (Configure) > ALARM > ALARM TRIG MODE > SINGLE|CONTIN
Remote Interface Operation:
The following program segment configures the
instrument to continuously scan when an alarm is detected.
Note:
To stop a scan, send the
ABORt
command.
TRIG:SOURCE ALARM1
TRIG:SOURCE:ALARM CONT
CALC:LIM:UPPER 10.25,(@1003)
CALC:LIM:UPPER:STATE ON,(@1003)
OUTPUT:ALARM1:SOURCE (@1003)
ROUT:MON:CHAN (@1003)
ROUT:MON:CHAN:ENABLE ON,(@1003)
ROUT:MON:STATE ON
INIT
Select alarm configuration
Select continuous scan mode
Set upper alarm limit
Enable alarms
Report alarms on Alarm 1
Select monitor channel
Enable monitoring on channel
Enable monitor mode
Initiate the scan
Summary of Contents for Agilent 34980A
Page 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Page 2: ......
Page 6: ......
Page 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Page 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Page 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Page 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Page 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...