4-83
Screens
Output RF Spectrum, Main View (Option 006 Only)
4. Mode
This field selects the Output RF Spectrum measurement Mode.
Choices
Ramping measures the Output RF Spectrum power due to
ramping. (The peak value is returned within the time interval 28
µ
s
before bit 0 to 28
µ
s after bit 147 or bit 87, depending on the burst
type.)
Ramp Ref makes a reference measurement needed for the Output
RF Spectrum measurement when making ramping measurements.
Modulatn (modulation) measures the Output RF Spectrum power
due to modulation.
Mod Ref makes a reference measurement needed for the Output
RF Spectrum measurement when making modulation
measurements.
5. Output RF
Spectrum
This field displays the measured Output RF Spectrum power at the specified Freq
Offset setting.
See Also
Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG, Units
6. Single/Cont
Single allows one test to be performed. Cont allows testing to automatically repeat.
7. Trig Delay
This field sets the time delay between a valid trigger event and the beginning of a
measurement.
Range
0 to 5000.00
µ
s
8. View
This field selects the alternate views of the Output RF Spectrum measurement.
Choices
Main displays the Output RF spectrum display as a text screen.
Trace displays Output RF Spectrum power spectral density (at the
Freq Offset setting) versus time.
Summary of Contents for 8922M
Page 8: ...Contents Contents 6...
Page 21: ...1 1 1 Installing Your Agilent 8922M S...
Page 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Page 29: ...2 1 2 Making Measurements...
Page 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Page 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Page 75: ...2 47 Making Measurements Advanced Features...
Page 76: ...3 1 3 Verifying Performance...
Page 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Page 236: ...4 130 Screens Tests...
Page 237: ...5 1 5 Keys...
Page 238: ...5 2 Keys Key Map Key Map...
Page 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Page 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Page 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Page 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Page 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Page 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Page 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Page 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Page 297: ...8 1 8 Instrument BASIC...
Page 341: ...A 1 A APPENDIX A...
Page 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Page 365: ...B 1 B Glossary...
Page 370: ...B 6 Glossary...