2-32
Making Measurements
If You Have Problems with a Measurement
Possible Solutions to
Message Line Errors
The signal processing hardware that is used to generate measurement results has to
be told when to take data samples. If the trigger to do this is misaligned with the
signal, there may be errors in the results that are displayed.
Trigger Timing A
Check if the trigger is being received.
On the MEAS SYNC screen (press
to gain access), check if the
correct burst type has been defined.
Trigger Timing B
Check if the trigger is being received at the correct time.
The First Bit field (
A
) , see Figure 2-17 on page 2-32, on the Phase Freq: Data
Bits screen displays the time difference between when a trigger is being received
and when the first bit of a burst occurred. The time difference is only valid if
FMErrCount is 0. The timing difference can be corrected by altering the trigger
delay (
B
).
Figure 2-17
MEAS SYNC
A
B
Summary of Contents for 8922M
Page 8: ...Contents Contents 6...
Page 21: ...1 1 1 Installing Your Agilent 8922M S...
Page 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Page 29: ...2 1 2 Making Measurements...
Page 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Page 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Page 75: ...2 47 Making Measurements Advanced Features...
Page 76: ...3 1 3 Verifying Performance...
Page 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Page 236: ...4 130 Screens Tests...
Page 237: ...5 1 5 Keys...
Page 238: ...5 2 Keys Key Map Key Map...
Page 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Page 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Page 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Page 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Page 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Page 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Page 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Page 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Page 297: ...8 1 8 Instrument BASIC...
Page 341: ...A 1 A APPENDIX A...
Page 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Page 365: ...B 1 B Glossary...
Page 370: ...B 6 Glossary...