4-58
Screens
Fast Bit Error
10. MS Loopback
This field is used to enter or display the Round Trip Delay (RTD) of the mobile
(expressed in number of TDMA frames), before bits are compared. The correct
number of TDMA frames can be determined automatically by making a Bit Error
Test using Auto Mode when the bit error ratio is approximately <20%. Then, if
faster measurements are desired or if the bit error ratio exceeds 20%, change the
mode to manual.
Choices
Burst Delay - range 0 through 26.
Manual
Auto
When Compatibility Switch mode is used (refer to Chapter 3,
Using Compatibility Switch), two additional fields appear under
the new heading
Loopback
.
Bit Error measurements are not automatic when in
Compatibility Switch mode. The additional fields allow manual
control of test loop commands to the mobile. When selecting
Off
, an open test loop command is sent to the mobile and
selecting
FAST
sends a close test loop command.
See Also
Screens: Bit Error Test (Manual, Auto)
11. Run/Stop
(Control)
Run starts a bit error test. Stop discontinues the bit error test in progress.
12. Single/Cont
(Control)
Single allows one test to be performed. Cont allows testing to automatically repeat.
13. % BE Ratio
This field allows you to select between a display showing the percentage of errors
(select
MHz
) and ppm (select
W
) which gives you the errors in parts-per-million.
See Also
Keys: ON/OFF, LO LIMIT, HI LIMIT, REF SET, AVG, Units
Summary of Contents for 8922M
Page 8: ...Contents Contents 6...
Page 21: ...1 1 1 Installing Your Agilent 8922M S...
Page 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Page 29: ...2 1 2 Making Measurements...
Page 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Page 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Page 75: ...2 47 Making Measurements Advanced Features...
Page 76: ...3 1 3 Verifying Performance...
Page 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Page 236: ...4 130 Screens Tests...
Page 237: ...5 1 5 Keys...
Page 238: ...5 2 Keys Key Map Key Map...
Page 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Page 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Page 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Page 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Page 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Page 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Page 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Page 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Page 297: ...8 1 8 Instrument BASIC...
Page 341: ...A 1 A APPENDIX A...
Page 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Page 365: ...B 1 B Glossary...
Page 370: ...B 6 Glossary...