3-15
Verifying Performance
Agilent Technologies 8922M/S Specifications
Output
Supplemental Characteristics:
Frequency Overrange
To 1015 MHz with uncalibrated output and modulation
Switching Speed
577
µ
s over the GSM frequency bands in hop mode (refer to 0.3
GMSK modulation specs
RF In/Out Connector
Level Range
−
14 to
−
127 dBm
Level Resolution
0.1 dB
Level Accuracy
2
GSM Bands
1
±
1.0 dB, levels
≥−
127dBm,
maximum over range >
−
12 dB
±
1.0 dB,
typically for levels
≥−
127 dBm while hopping.
50 MHz to 1 GHz
±
1.5 dB, levels
≥−
107 dBm
±
2.0 dB, levels
≥−
127 dBm
10 MHz to 50 MHz
±
2.0 dB, levels
≥−
107 dBm
±
2.5 dB, levels
≥−
127 dBm
Reverse Power
15 Watts continuous
100 Watts for 10 seconds/minute
SWR
1.5:1
Aux RF Out Connector
Level Range
+4 to
−
127 dBm
Level Resolution
0.1 dB
Level Accuracy
2
GSM Bands
1
±
1.0 dB, levels
≥−
107 dBm
±
1.0 dB, typically for levels
≥−
107 dBm while hopping.
50 MHz to 1 GHz
±
1.5 dB, levels
≥−
107 dBm
±
2.0 dB, levels
≥−
127 dBm
10 MHz to 50 MHz
±
2.0 dB, levels
≥−
107 dBm
±
2.5 dB, levels
≥−
127 dBm
Reverse Power
200 milliwatts
SWR
2.0:1, level <
−
4 dBm
1. GSM frequency bands are 880 to 915 MHz and 925 to 960 Mhz.
2. Level accuracy degrades 0.2 dB when using the RF in/out connector for both RF generator and RF analyzer. In 30dB
pulse mode, level accuracy specifications are typical.
Summary of Contents for 8922M
Page 8: ...Contents Contents 6...
Page 21: ...1 1 1 Installing Your Agilent 8922M S...
Page 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Page 29: ...2 1 2 Making Measurements...
Page 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Page 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Page 75: ...2 47 Making Measurements Advanced Features...
Page 76: ...3 1 3 Verifying Performance...
Page 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Page 236: ...4 130 Screens Tests...
Page 237: ...5 1 5 Keys...
Page 238: ...5 2 Keys Key Map Key Map...
Page 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Page 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Page 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Page 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Page 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Page 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Page 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Page 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Page 297: ...8 1 8 Instrument BASIC...
Page 341: ...A 1 A APPENDIX A...
Page 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Page 365: ...B 1 B Glossary...
Page 370: ...B 6 Glossary...