2-41
Making Measurements
Advanced Features
Power Ramp Setting
Time
The mobile transmitter output ramp should settle so that it complies with the power
mask (GSM Rec 11.10) by the time output level calibration is done in the
manufacturing process. The power mask is displayed in the Power Ramp screens.
Peak Carrier Power measurement method is not made to GSM Rec. 11.10. If you
need to make a measurement to this specification, go to the Power Ramp Summary
screen. The field in the top right-hand corner displays the Peak Carrier Power to
GSM Rec. 11.10. Refer to PWR RAMP: SUMMARY screen (item 5) for a
description.
Peak Carrier Power
Measurement,
Assumptions And
Algorithm
The algorithm for the Peak Carrier Power measurement is based on the use of a fast
sampler taking continuous asynchronous measurements of the power detected at the
RF IN/OUT port.
The sampler accumulates 75
µ
S sample bins containing the maximum power
observed during the period and the minimum power observed during the period.
Sampling rate is 10 MHz. The maximum and minimum values observed are based
on 750 individual samples within the period.
A trigger is provided at the middle of bit 0 of the burst. (When using RF Rise
triggering, the actual trigger may occur several bits earlier due to the ramp up of the
burst.) In response to this trigger, the following 6 sample bins (0 - 5) from the
sampler are saved and reported to the measurement processor.
-2
-1
0
1
2
3
4
5
6
7
8
9
0
µ
S
577
µ
S
Timeslot
Mobile Burst
Amplitude
Trigger
Sample
75
µ
S/sample_bin
375
µ
S Window
Bins
Summary of Contents for 8922M
Page 8: ...Contents Contents 6...
Page 21: ...1 1 1 Installing Your Agilent 8922M S...
Page 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Page 29: ...2 1 2 Making Measurements...
Page 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Page 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Page 75: ...2 47 Making Measurements Advanced Features...
Page 76: ...3 1 3 Verifying Performance...
Page 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Page 236: ...4 130 Screens Tests...
Page 237: ...5 1 5 Keys...
Page 238: ...5 2 Keys Key Map Key Map...
Page 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Page 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Page 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Page 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Page 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Page 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Page 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Page 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Page 297: ...8 1 8 Instrument BASIC...
Page 341: ...A 1 A APPENDIX A...
Page 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Page 365: ...B 1 B Glossary...
Page 370: ...B 6 Glossary...