3-7
Verifying Performance
Understanding the Tests
Understanding the Tests
This section describes the theory of each performance test, lists the equipment
needed for the test, and provides some problem solving information.
Test 01: Signal
Generator Level
Equipment Required
Measuring Receiver HP/Agilent 8902A
Sensor Module HP/Agilent 11722A
Theory of the Test
The UUT is set to generate CW signals at various levels and frequencies. The HP/
Agilent 8902A/11722A is used as a power meter to measure high level signals and
the HP/Agilent 8902A tuned RF level function is used to measure low level signals.
Things To Check In Case Of Problems
•
The HP/Agilent 8902A/11722A must be calibrated and zeroed for accurate
measurement results. The HP/Agilent 11722A calibration factors must be entered
into the HP/Agilent 8902A.
•
Low level signals from the Agilent 8922M/S must be very stable to be measured
by the tuned RF level function of the HP/Agilent 8902A. The instrument
timebases should be locked together. Also, if the Agilent 8922 source has
unusually high residual FM it may cause the low level measurements to fail
because the HP/Agilent 8902A cannot lock onto the signal.
•
Mismatch uncertainty causes a high level of uncertainty in this measurement.
Summary of Contents for 8922M
Page 8: ...Contents Contents 6...
Page 21: ...1 1 1 Installing Your Agilent 8922M S...
Page 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Page 29: ...2 1 2 Making Measurements...
Page 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Page 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Page 75: ...2 47 Making Measurements Advanced Features...
Page 76: ...3 1 3 Verifying Performance...
Page 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Page 236: ...4 130 Screens Tests...
Page 237: ...5 1 5 Keys...
Page 238: ...5 2 Keys Key Map Key Map...
Page 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Page 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Page 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Page 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Page 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Page 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Page 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Page 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Page 297: ...8 1 8 Instrument BASIC...
Page 341: ...A 1 A APPENDIX A...
Page 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Page 365: ...B 1 B Glossary...
Page 370: ...B 6 Glossary...