4-43
Screens
Cell Control 2
25. RACHs
This field indicates the number of RACHs received during the current call or since
the last Reset.
26. Relative MS
Timing Err
This field displays the length of time between when the Agilent 8922M/S expected
the uplink burst to arrive and the time it actually arrived.
Timing error measurements on the uplink burst are made from the center of bit 0,
and are relative to the default trigger delay value of 473.4 T (see Measurement Sync,
Trigger Delay).
The Agilent 8922M/S normally expects the uplink burst 468.75 bit periods after the
beginning of the downlink burst. The trigger delay’s default value of 473.4 T reflects
an additional Agilent 8922M/S delay of 4.65 T (468.75 + 4.65 = 473.4 T).
The accuracy of this measurement is
±
0.25 T. For more accurate measurements,
refer to DSP Analyzer, Data Bits, First Bit.
See Also
Screens: Phase/Freq: Data Bits (
First Bit
)
27. Reset (Call
Counts)
This field clears out the counts of pages, RACHs, synchronization, and decode
errors detected.
28. Reset (SACCH
Meas)
This field clears-out all of the SACCH measurements.
Summary of Contents for 8922M
Page 8: ...Contents Contents 6...
Page 21: ...1 1 1 Installing Your Agilent 8922M S...
Page 24: ...1 4 Installing Your Agilent 8922M S Fuses and Power Cords Power Cords Agilent Part Agilent...
Page 29: ...2 1 2 Making Measurements...
Page 59: ...2 31 Making Measurements If You Have Problems with a Measurement Figure 2 16...
Page 61: ...2 33 Making Measurements If You Have Problems with a Measurement Trigger Range For Pulsed RF...
Page 75: ...2 47 Making Measurements Advanced Features...
Page 76: ...3 1 3 Verifying Performance...
Page 106: ...3 31 Verifying Performance Agilent Technologies 8922M S Specifications...
Page 236: ...4 130 Screens Tests...
Page 237: ...5 1 5 Keys...
Page 238: ...5 2 Keys Key Map Key Map...
Page 276: ...6 28 Connectors Timing Diagrams Figure 6 3 Digital Demod Timing Diagram...
Page 277: ...6 29 Connectors Timing Diagrams Figure 6 4 Frequency Hop Timing Specification Table...
Page 278: ...6 30 Connectors Timing Diagrams Figure 6 5 RF Generator Hop Address Source Ext Timing Diagram...
Page 279: ...6 31 Connectors Timing Diagrams Figure 6 6 RF Analyzer Hop Address Source Ext Timing Diagram...
Page 280: ...6 32 Connectors Timing Diagrams Figure 6 7 RF Generator Hop Address Source Int Timing Diagram...
Page 281: ...6 33 Connectors Timing Diagrams Figure 6 8 RF Analyzer Hop Address Int Timing Diagram...
Page 282: ...6 34 Connectors Timing Diagrams Figure 6 9 Reset Hop Once Address Source Int Timing Diagram...
Page 284: ...6 36 Connectors Timing Diagrams Figure 6 11 Hop Inhibit Address Source Int Timing Diagram...
Page 297: ...8 1 8 Instrument BASIC...
Page 341: ...A 1 A APPENDIX A...
Page 364: ...A 24 APPENDIX A Protocol Log of a Typical Call...
Page 365: ...B 1 B Glossary...
Page 370: ...B 6 Glossary...