![Agilent Technologies 86120C User Manual Download Page 89](http://html.mh-extra.com/html/agilent-technologies/86120c/86120c_user-manual_2867947089.webp)
3-11
Programming
Making Measurements
Also, because new data is not collected, FETCh is especially useful
when characterizing transient data.
FETCh does not reconfigure the disp lay. For examp le, if the disp lay is
in the
Peak WL
mode, sending :FETCh:ARRay does not configure the
display to the
List by WL
even though an array of data is returned to
the computer.
A common p rogramming error occurs when the :FETCh command is
used after an *RST command. This generates error number –230,
“Data corrupt or stale”
. In this instance, you must send :INIT:IMM
after the *RST command and before the :FETCh command to cap ture a
new array of measurement data.
:CONFigure command
The CONFigure command changes measurement settings without taking
a measurement. The instrument is p laced in the
List by WL
,
List by
Ampl
, or
Peak WL
display application.
CONFigure can be queried. The query returns the last configuration
setup by the CONFigure command. The instrument returns a string
which is the last instrument function sent by a CONFigure command
or MEASure query. The returned string is in the short command form.
Use caution when using this query, because if any instrument settings
were changed since the last CONFigure command or MEASure query
these changes may not be included in the returned string.
For examp le, if the last CONFigure command was:
:CONFigure:SCALar:POWer:WAVelength 1300NM, MAX
a CONFigure? query would return a string that is similar to the follow-
ing line:
“POW:WAV 1.300000e-6,0.01”
The
1300NM
and
resolution
values track the actual instrument settings
and input signals. Notice that the quotation marks are part of the
returned string.
Return single or multiple measurement values
You can specify whether FETCh, READ, or MEASure returns a single
value (SCALar) or multiple values (ARRay). The following example
specifies SCALar data which returns a single value.
:MEASure:SCALar:POWer:WAVelength? MAX
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......