4-98
Programming Commands
SYSTem Subsystem
ERRor
Queries an error from the error queue.
Syntax
:SYSTem:ERRor?
Attribute
Summary
Preset
State: none
*RST State: none
SCPI Compliance: standard
Query Only
Description
The Agilent 86120C has a 30 entry error queue. The queue is a first-
in, first- out buffer. Repeatedly sending the query :SYSTEM:ERROR?
returns the error numbers and descriptions in the order in which they
occur until the queue is empty. Any further queries returns
+0, “No
errors”
until another error occurs.
For a complete list of error messages,
refer to “Error Messages” on
page 7- 11
.
Query Response
<value>, <string>
<value>
is an integer.
<string>
is the text of the error message. The
following is an example of a response:
-113,”Undefined header”
Example
DIM Error$[250]
OUTPUT 720;”:SYSTEM:ERROR?”
ENTER 720;Error$
PRINT Error$
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......