2-31
Making Measurements
Measuring Fabry-Perot (FP) Lasers
Measuring Fabry- Perot (FP) Lasers
The Agilent 86120C can perform several measurements on Fabry- Perot
lasers including FWHM and mode sp acing. The disp lay shows the mea-
surement results in the selected wavelength and amplitude units. In
addition, the mode spacing measurement always shows results in fre-
quency as well as the selected wavelength units.
Refer to “Displayed
units” on page 2- 12
to learn how to change the units. The number of
laser lines included in the measurement results is also listed as shown
in the following figure.
To characterize a Fabry- Perot laser
1
Press
Appl’s
.
2
Press
FP TEST
to measure the Fabry- Perot laser’s characteristics.
If you want to stop the up dating of measurement data with each
sweep, press
Single
. Because Fabry- Perot lasers are sensitive to reflec-
tions it is good measurement p ractice to p lace an op tical isolator or
attenuator between the laser and the Agilent 86120C.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......