2-26
Making Measurements
Measuring Signal-to-Noise Ratios
Location of noise measurements
Automatic
interpolation
When the signal- to- noise “auto” function is selected, the
Agilent 86120C first determines the p roximity of any adjacent signal. If
the next closest signal is
≤
200 GHz (approximately 1.6 nm at 1550 nm)
away from the signal of interest, then the noise p ower is measured
half way between the two channels and an equal distance to the other
side of the signal of interest. See points P
n1
and P
n2
in the following
figure.
If the closest signal is more than 200 GHz from the signal of interest,
or if there is no other signals p resent, then the noise p ower is mea-
sured at 100 GHz on either side of the signal of interest. The two mea-
sured noise power levels are then averaged to estimate the noise
p ower level at the signal wavelength. The noise p ower measurements
use linear interpolation to estimate the noise power level at the signal
of interest’s wavelength.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......