3-4
Programming
Addressing and Initializing the Instrument
Notice in the examp le above, that the commands are sent to an instru-
ment address of 720. This indicates address 20 on an interface with
select code 7. Pressing the green
Preset
key does not change the GPIB
address.
Set single acquisition mode
An advantage of using the *RST command is that it sets the
Agilent 86120C into the single measurement acquisition mode. Because
the READ and MEASure data queries expect this mode, their proper
operation is ensured.
To change the GPIB address
1
Press the
Setup
key.
2
Press
MORE
twice, then
GPIB
.
3
Use the
and
softkeys to change the GPIB address.
4
Press
RETURN
.