6-8
Specifications and Regulatory Information
Specifications—FAST Update Mode
Specifications—FAST Update Mode
Each laser line is assumed to have a linewidth (including modulation side-
bands) of less than 10 GHz.
All specifications apply when the instrument is in the following modes:
•
FAST
up date mode unless noted.
Refer to “Measurement rate” on
page 2- 13
.
• Configured to measure narrowband devices. Sp ecifications do
not
ap ply
when the instrument is configured to measure broadband devices.
Refer
to “Measuring broadband devices and chirp ed lasers” on p age 2- 9
.
Wavelength
Range
1270–1650
Absolute accuracy (lines separated by
≥
30 GHz)
±
3 ppm (
±
0.005 nm at 1550 nm,
±
0.004 nm at 1310 nm)
Differential accuracy (characteristic)
±
2 ppm
Minimum resolvable separation
a
(characteristic)
20 GHz (0.16 nm at 1550 nm, 0.11 nm at 1300 nm)
Display resolution
0.001 nm
a. Signals of equal amplitude. For lines that are separated by less than 30 GHz, wavelength accuracy is reduced.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......