viii
General Safety Considerations
C A U T I O N
This product complies with Overvoltage Category II and Pollution Degree 2.
C A U T I O N
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4
°
C for every 100 watts dissipated in the cabinet.
If the total power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
C A U T I O N
Always use the three-prong ac power cord supplied with this instrument. Failure
to ensure adequate earth grounding by not using this cord may cause
instrument damage.
C A U T I O N
Do not
connect ac power until you have verified the line voltage is correct as
described in
“Line Power Requirements” on page 1-5
. Damage to the equipment
could result.
C A U T I O N
This instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......