4-51
Programming Commands
CALCulate3 Subsystem
DELTa:REFerence:FREQuency
Selects the reference laser line for DELTa calculations.
Syntax
:CALCulate3:DELTa:REFerence:FREQuency{?| {<real> | MINimum | MAXimum}}
<real>
is a frequency value that is within the following limits:
Constant
Description
MINimum 181.6924
THz
MAXimum 236.0571
THz
Attribute
Summary
Preset
State: 236.0571 THz (1270 nm)
*RST State: 236.0571 THz (1270 nm)
SCPI Comp liance: instrument sp ecific
Description
The reference will be the laser line at the frequency closest to the fre-
quency entered. Subsequent measurements will use the frequency clos-
est to the reference frequency used for the previous measurement.
The query returns the reference laser line’s frequency. The default
units for the
<real>
p arameter are Hz.
DELTa:REFerence:POWer?
Queries the reference laser line’s power level.
Syntax
:CALCulate3:DELTa:REFerence:POWer?
Attribute
Summary
Preset
State: not affected
*RST State: not affected
SCPI Comp liance: instrument sp ecific
Query Only
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......