![Agilent Technologies 86120C User Manual Download Page 150](http://html.mh-extra.com/html/agilent-technologies/86120c/86120c_user-manual_2867947150.webp)
4-22
Programming Commands
Measurement Instructions
<resolution>
Constants
MAXimum
0.01 resolution (fast update)
MINimum
0.001 resolution (normal)
DEFault Current
resolution
Examples
:CONF:ARR:POW:WAV DEF, MAX
:FETC:ARR:POW:WAV? DEF, MIN
:READ:ARR:POW:WAV?
:MEAS:ARR:POW:WAV?
:CONF:SCAL:POW:WAV 1300NM, MAX
:FETC:SCAL:POW:WAV? 1300NM, MIN
:READ:SCAL:POW:WAV? 1300NM
:MEAS:SCAL:POW:WAV? 1300NM
Query Response
The following line is an examp le of a returned string when
:MEAS:SCAL:POW:WAV? MAX is sent:
+1.5529258E-006
If six laser lines are located and :MEAS:ARR:POW:WAV? is sent, the
following string could be returned. Notice that the first returned num-
ber indicates the number of laser- line values returned in the query.
6,+1.54488100E-006,+1.54648400E-006,+1.54809000E-006,+1.54969900E-006,+1.55131100E-
006,+1.55292600E-006
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......