2-4
Making Measurements
Measuring Wavelength and Power
Peak WL mode
When
Peak WL
is pressed, the display shows the largest amplitude line
in the sp ectrum. The word
PEAK
is shown on the screen. If multip le
laser lines are present at the input, the number of lines located will be
shown along the right side of the screen. In peak wavelength mode, the
Agilent 86120C can measure up to 200 laser lines simultaneously.
Figure 2-1. Display after “Peak WL” key pressed
In addition to the digital readouts, there is a p ower bar. It p rovides a
convenient analog “meter movement” for tuning laser p ower.
Although the
Peak WL
mode shows one signal at a time, softkeys are
provided that allow you to scroll through and display all the measured
laser lines. You can scroll through the list according to the wave-
lengths or powers measured.
To display peak wavelength and power
1
Connect the fiber- optic cable to the front- panel
OPTICAL INPUT
connector.
2
To disp lay the p eak wavelength and p ower, do one of the following:
• Press the green
Preset
key.
• Press
Peak WL
.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......