3-6
Programming
Making Measurements
After collecting the uncorrected data, the Agilent 86120C searches the
data for the first 200 p eak resp onses. (For WLIMit:OFF, searching
starts at 1650 nm and progresses towards 1270 nm. For WLIMit:ON,
searching starts at WLIMit:START and p rogresses toward
WLIMit:STOP.) These p eak values are then p laced into the corrected
data buffer. Each p eak value consists of an amp litude and wavelength
measurement. Amplitude and wavelength correction factors are applied
to this data.
For a listing of the p rogramming commands (including a cross refer-
ence to front- p anel keys), refer to the following tables:
Table 3- 10, “Programming Commands,” on page 3- 43
Table 3- 11, “Keys Versus Commands,” on page 3- 48