2-19
Making Measurements
Measuring Laser Separation
Channel separation
Sup p ose that you want to measure sep aration on a system having the
sp ectrum shown in the following figure.
The Agilent 86120C displays separation on this spectrum as shown in
the following figure. Notice that the 1541.747 nm laser line is selected
as the reference. It is shown in absolute units. The wavelengths and
powers of the remaining responses are shown relative to this refer-
ence. For example, the first response is 2.596 nm below the reference.
To determine channel spacing, simply read the relative wavelength
measurement of the laser lines immediately p receding and following
the reference. Use the
,
, and
SELECT
softkeys to change the ref-
erence laser line and read the channel sp acing between each channel.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......