![Agilent Technologies 86120C User Manual Download Page 187](http://html.mh-extra.com/html/agilent-technologies/86120c/86120c_user-manual_2867947187.webp)
4-59
Programming Commands
CALCulate3 Subsystem
DRIFt:PRESet
Turns off all the drift states for DIFFerence, MAXimum, MINimum, and
REFerence.
Syntax
:CALCulate3:DRIFt:PRESet
Attribute
Summary
Preset
State: unaffected by
*RST State: unaffected by
SCPI Comp liance: instrument sp ecific
Command Only
Description
This command allows the CALC3:DATA? query to return the difference
between the current measurement and the reference.
DRIFt:REFerence:RESet
Places the current list of laser lines into the reference list.
Syntax
:CALCulate3:DRIFt:REFerence:RESet
Attribute
Summary
Preset
State: unaffected by
*RST State: unaffected by
SCPI Comp liance: instrument sp ecific
Command Only
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......