4-34
Programming Commands
CALCulate2 Subsystem
POINts?
Queries the number of points in the data set.
Syntax
:CALCulate2:POINts?
Attribute
Summary
Preset
State: unaffected
*RST State: unaffected
SCPI Comp liance: instrument sp ecific
Query Only
Description
This is the number of p oints that will be returned by the
CALC2:DATA? query.
Query Response
For examp le, if six laser lines are located:
+6
PTHReshold
Sets the p eak threshold limit used by the instrument to determine
valid laser line peaks.
Syntax
:CALCulate2:PTHReshold{?| {<integer> | MINimum | MAXimum | DEFault}}
<integer>
rep resents logarithmic units in dB. Valid range is 0 to 40.
Constant
Value
MINimum 0
dB
MAXimum 40
dB
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
Refer
to “Always force the Agilent 86120C to wait for non-sequential commands” on
page 3-12
for more information.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......