Upper Power Limit and Resolution
Table 9 shows the upper power limit and measurent resolution at various
power ranges. As can be seen the resolution decreases as the chosen
[Range] decreases. The resolution is always 40 dBm less than the chosen
[Range] value. The Upper Power Limit is always 3 dBm greater than the
chosen [Range] value.
Table 9 Upper Power Limits and Resolution for Various Power Ranges
[R
ANGE
]
Upper Power Limit
Resolution
40 dBm
19.999 W
43.000 dBm
0.001 W
−
0.000 dBm
30 dBm
1.9999 W
33.000 dBm
0.1 mW
−
10.000 dBm
20 dBm
199.99 mW
23.000 dBm
0.01 mW
−
20.000 dBm
10 dBm
19.999 mW
13.000 dBm
0.001 mW
−
30.000 dBm
0 dBm
1.9999 mW
3.000 dBm
0.1
µ
W
−
40.000 dBm
−
10 dBm
199.99
µ
W
−
7.000 dBm
0.01
µ
W
−
50.000 dBm
−
20 dBm
19.999
µ
W
−
17.000 dBm
0.001
µ
W
−
60.000 dBm
−
30 dBm
1.9999
µ
W
−
27.000 dBm
0.1 nw
−
70.000 dBm
−
40 dBm
199.99 nW
−
37.000 dBm
0.01 nw
−
80.000 dBm
−
50 dBm
19.999 nW
−
47.000 dBm
0.001 nw
−
90.000 dBm
−
60 dBm
1.9999 nW
−
57.000 dBm
0.1 pW
−
100.000 dBm
−
70 dBm
199.99 pW
−
67.000 dBm
0.01 pW
−
110.000 dBm
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......