DFB Laser Sources
Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
413
error message
E10060: Laser <a> control failed (b, c, d, e)
description
Checks the laser current source
test value
a
laser number
b
measured current at low per and normal current source
setting
c
measured current at low power and high current source
setting
d
measured current at high power and normal current
source setting
e
measured current at high power and high current source
setting
action
Please return the unit to your local service center for repair
error message
E10080: Laser <a> control for safety test failed
description
To test the current safety comparator the laser current needs to
be set to two different values
test value
a
laser number
action
Please return the unit to your local service center for repair
error message
E10081: Safety comparator <a> test failed (b, c)
description
Checks whether laser current safety comparator includes at the
right conditions a safety interrupt at the CPU
test value
a
laser number
b
shows whether the interrupt flag was raised when the
current was lower than the limit
c
shows whether the interrupt flag was raised when the
current was higher than the limit
action
Please return the unit to your local service center for repair
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......