How to Use a Variable Optical Attenuator module
Setting Attenuation and/or Power Levels
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
213
Figure 125 The
λ
Offset Table
Figure 125 shows the fourth, fifth and sixth entries in a
λ
Offset Table
displayed on an 8164A/B mainframe.
•
To display the previous three entries, press the [Prev] softkey.
•
To display the next three entries, press the [Next] softkey.
4
Press the [Add] softkey.
•
Enter the wavelength
λ.
•
Select, then enter the corresponding Offset
5
Press [OK] to store the result in the
λ
Offset Table.
• The
λ
Offset Table is automatically ordered by wavelength, there is no need to
enter wavelenght:offset pairs in any particular order.
• To edit the offset applied to an existing entry, use the cursor key to select its
λ
field, press [Add], enter the new offset, then press [OK].
• To remove an existing entry, use the cursor key to select its
λ
field, then press
[Del]. The indices of the remaining table entries are automatically updated;
[Del] does not leave a blank entry.