Applications
The Logging Application
242
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
The Logging Application
The Logging data acquisition application logs a series of power
measurements for a number of Power Meter channels, plots the results as
a graph, and generates a statistical analysis of the results.
Figure 145 Example Logging Application
The most important parameters for the logging application are:
• [Samples], this is the number of samples that will be performed during the
logging application, and
• [AvgTime], the averaging time is the length of time over which a signal is
averaged. Longer averaging times increase the accuracy and improve the noise
rejection. Longer averaging times also decrease sensitivity.
The total time for a logging application is the number of samples multiplied
by the averaging time.
The Logging application differs from the “
The Stability Application
” on
page 254 because all power measurements are performed consecutively
without the possibility of any pause between measurements, as shown in
Figure 145 .
All results can be:
• displayed on the screen,
• printed out to hardcopy, or
• saved to disk drive of your Agilent 8164A/B Lightwave Measurement System.
Averaging Time
t
1
3
5
2
4
6
8
7
9
t
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......