Firmware Upgrades
Firmware Upgrade Process
458
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Agilent 8164B Firmware Upgrade Procedure
To upgrade the 8164B firmware:
1
Copy the Firmware\8164B folder from the Support CD-ROM to a folder on your
PC. This folder contains the update tool ATTools_v1_23.exe and a firmware
image file named 64B_Vxxx.Z
“xxx” in a filename refers to the firmware version. For example, 64B_V402.Z is
firmware version 4.02 for the 8164B instrument.
2
Connect your instrument to your PC with the serial cable.
3
Turn on the 8164B and wait until the Agilent screen appears.
Quickly:
•
Press, but do not hold the Aux hardkey
•
Press the Appl hardkey.
If you were successful, the instrument displays:
“Instrument Maintenance Mode”.
Otherwise, turn off the power and try again.
4
At your PC, click on ATTools_v1_23.exe to start the update tool.
Click OK to confirm that you have completed steps 1 to 3.
When the serial connection has been established, which may be after a pause
of around 30 seconds, the ATTools user interface is displayed.
5
Select the firmware image (file type .Z) in the file selection box, then click the
Update Firmware button.
Click OK to confirm your choice of firmware upgrade.
6
The firmware upgrade is tracked by a progress bar in the the ATTools user
interface.
When the update is complete, a confirmation message is displayed.
Click OK
7
Click Exit to terminate ATTools.
Your 8164B reboots automatically.
The upgrade process with floppy disks used for the 8164A is not supported
for 8164B instruments.
You require a PC, running MS Windows 95, NT 4.0, or higher,
and a null-modem serial cable, such as is supplied with your 8164B
mainframe.
N O T E
Summary of Contents for 8163B
Page 4: ......
Page 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Page 465: ......